Exiting The Calibration Mode - A&D AD-4405A Instruction Manual

Weighing indicator
Table of Contents

Advertisement

Weighing a Mass less than the Maximum Capacity
Step12 Set a mass value using the
Step13 Place a mass equivalent to the displayed value
on the weighing unit. Proceed to step 15.
Weighing a Maximum Capacity Mass
Step14 Place a mass equivalent to the maximum
capacity on the weighing unit.
Step15 Wait for the STABLE mark to turn on and
press the
key .......... The key to proceed to step17
ESC
Caution: Do not press the
STABLE mark is off (detecting motion).
Arrange the condition using the F00 filter
function.
Digital Input
Step16 Press the
parameter of the span in the unit of mV/V.
Adjust the input voltage using the
(It is possible to store a greater value than
the capacity.)
key ..... The key to store the parameter
ENTER
key .......... The key to proceed to step 17
ESC
5.2.5.

Exiting the Calibration Mode

Step17 Calend is displayed.
Use the following keys.
key ..... To store the parameters and display
CAL
key ..... The key to store the parameters
ESC
5.2. Calibration Procedure
0
key. Proceed to step17.
ENTER
without changing the parameter.
ENTER
key to display the input voltage
+/-
and proceed to step 17.
without changing the parameter.
Caloff. Proceed to the next step.
Caloff
can be displayed by pressing
the
key while pressing the
+/-
instead of
key.
CAL
temporarily. Proceed to the
display.
-
keys.
9
key while the
-
keys.
0
9
key
ON/OFF
Cal 0
Page 16
Span
Maximum
To step 14, step 16
New mass value
Mass of
maximum capacity
To step 14
To step 16
Turn on STABLE mark
NET/GROSS
Stored input voltage
+/-
mV/V
0
9
mV/V
PRINT
New input voltage
ENTER
or
PRINT
ENTER
Turn display off
AD-4405A Weighing Indicator
To step 12
0
9
step 13
step 14
Place the
displayed
mass on the
PRINT
ENTER
NET/GROSS
PRINT
+/-
ENTER