have been given names, these are shown when the process file is selected in
the list. Click 'No Process' if you do not want to apply a process file.
(A process file allows you to see the temperature profile in relation to the
furnace zones as the profile appears on screen during the run. See the
Insight software for an introduction to process files: press function key F1,
or select Help > Contents from the menu bar, and click the section
'Process Files: Furnace, Recipe, Product'.)
You may specify that a password is required when an attempt is made to
close Insight while a real-time telemetry run is in progress: select Tools >
Options > General from the menu bar.

Real-time Display During the Run

After the first few data packets have been received, the data starts to be
displayed in the Graph and Analysis Windows, scrolling in real time as new data
is received. You may change the way the data is displayed with the Axes tab of
the Graph Options dialog (from the right-click menu, or from the main menu
select View > Graph Options): under Telemetry, specify how much of the
recently received data is displayed, and whether you wish to see only a certain
temperature (y-axis) range, centered on a the latest data.
You may zoom the display as when viewing a paqfile (see the online Help
system), except that:
Double-clicking on the graph (or selecting Real Time Zoom from the View
menu or right-click menu) shows only the most recently received portion of
the data on the scrolling graph (see above).
Saved zoom modes are not available.
If the y-axis is not set to be centered (see above), the default y-axis zoom
changes as more data is received, in order to accommodate all received data.
To move the graph across the viewing area, hold Shift and drag the mouse
pointer.
You may overlay one or more tolerance/ideal curves on the graph to
compare with the data as it is being received (select View > Overlay) (not
available with Insight Furnace Surveying software). Other paqfiles cannot be
overlaid.
You may adjust the furnace start position while a real-time run is in progress
(select Process > Adjust Furnace Start, or use the right-click menu).
Calculations shown in the Analysis Window for the chosen data analysis
mode update continuously as new data is received. As for non-real-time runs,
calculations are performed only on the currently zoomed area shown on the
graph. However, if the graph is scrolling and showing just the most recently
DATAPAQ XDL12