Hach Programs—Simplified Analysis

3.5.5 Running a Standard Adjust

3.5.6 Performing Standard Additions

3.6 Adding a Hach Program to the "Favorite Programs" List

Hach Programs—Simplified Analysis
Page 36
The DR/2400 Procedures Manual indicates which procedures can use the Standard
Adjust and provides instructions for performing a Standard Adjust. The general
procedure is explained here.
Read a standard before setting Standard Adjust to on.
1. Follow the entire Hach procedure (as described in the
DR/2400 Procedures Manual), using a known standard for the
sample.
2. After reading the concentration, touch
.
Adjust
3. If Standard Adjust is set to off, turn it on.
4. The box will show the default standard value for the test, as
mentioned in the procedure. To enter a different standard value,
touch the box and enter the value. Touch
The Standard Adjust icon will appear on the measurement screen
when Standard Adjust is enabled.
Note: The adjustment must be within certain limits, which vary with each
program.
Running standard additions helps to confirm the accuracy of the measurements.
The DR/2400 Procedures Manual and Standard Additions on page 43 describe in
detail how to use this function.
It is possible to add a Hach program to the "Favorites" list that contains
frequently used programs:
1. Touch
Hach Programs
2. Scroll through the list of programs and highlight the program to be added, or
touch
Select By Number
3. Touch
Add to Favorites
4. Touch
.
OK
5. The program can now be selected from the list of
Main Menu.
in the Main Menu.
for a known program number.
.
, then
Options
Standard
.
OK
on the
Favorite Programs
5940018HachPrograms.fm