Siemens SITRANS LVL100 Operating Instructions Manual page 11

Contactless electronic switch
Table of Contents

Advertisement

Moisture
Transport
Pressure/Vacuum
1
2
Fig. 3: Horizontal mounting
1
Switching point
2
Switching point (recommended mounting position, particularly for adhesive
products)
Use the recommended cables (see chapter "Connecting to power
supply") and tighten the cable gland.
You can give your SITRANS LVL100 additional protection against
moisture penetration by leading the connection cable downward in
front of the cable gland. Rain and condensation water can thus drain
off. This applies mainly to outdoor mounting as well as installation in
areas where high humidity is expected (e.g. through cleaning pro-
cesses) or on cooled or heated vessels.
Fig. 4: Measures against moisture ingress
Do not hold SITRANS LVL100 on the tuning fork.
The process fitting must be sealed if there is gauge or low pressure in
the vessel. Before use, check if the seal material is resistant against
the measured product and the process temperature.
The max. permissible pressure is specified in chapter "Technical
data" or on the type label of the sensor.
SITRANS LVL100 - Operating Instructions