(4) Connecting to Pipe
Please pass bolts from pipe line side, not flow tube
side, and tighten the bolts according to Torque Values
in Table 3.3.9.
CAUTION
• Please be sure to tighten the bolts following
prescribed torque values. Please tighten the
flange bolts diagonally with the same torque
values, step by step up to the prescribed torque
value.
• Please completely cover the brackets and liners
with mortar.
Figure 3.3.7
Mounting Flow Tube
Table 3.3.13 Tightening Torgue Values for Metal Piping
Flange
JIS G3451 F12/JIS7.5K
Rating
SIze:
N-m
kgf-cm
in-Ibf
mm (inch)
121
1234
1071
500 (20)
to 150
to 1530
to 1327
116
1183
1027
600 (24)
to 145
to 1479
to 1283
173
1765
1531
700 (28)
to 215
to 2193
to 1903
180
1836
1593
800 (32)
to 224
to 2285
to 1982
102
1040
903
900 (36)
to 127
to 1295
to 1124
231
2356
2044
1000 (40)
to 288
to 2938
to 2549
298
3040
2637
1100 (44)
to 372
to 3794
to 3292
301
3070
2664
1200 (48)
to 376
to 3835
to 3327
432
4406
3823
1350 (54)
to 539
to 5498
to 4770
463
4723
4097
1500 (60)
to 578
to 5896
to 5115
468
4774
4142
1600 (64)
to 585
to 5967
to 5177
485
4947
4292
1800 (72)
to 605
to 6171
to 5354
615
6273
5442
2000 (80)
to 768
to 7834
to 6796
686
6997
6071
2200 (88)
to 858
to 8752
to 7593
759
7742
6717
2400 (96)
to 948
to 9670
to 8389
1006
10261
8903
2600 (104)
to 1257
to 12821
to 11124
* Please use rubber gasket or the gasket which has equal elasticity.
3.3.6
Mounting Procedure for Sani-
tary Type
The sanitary type is mounted using IDF clamps.
(1) Mounting Direction
Please mount the Magnetic Flow Tube matching the
flow direction of the fluid to be measured with the
direction of the arrow mark on the flow tube.
ANSI150
JIS10K
N-m
kgf-cm
in-Ibf
N-m
kgf-cm
63
643
558
59 to 73
602 to 745
to 79
to 806
to 699
78
796
690
67 to 84
683 to 857
to 97
to 989
to 858
104
1061
920
to 130
to 1326
to 1150
115
1173
1018
to 144
to 1469
to 1274
175
1785
1549
to 182
to 1856
to 1611
204
2081
1805
to 254
to 2591
to 2248
IMPORTANT
If it is impossible to macth the direction, please
never remodel to change direction of the termi-
nal box. In case the measuring fluid flows
against the arrow direction, please refer to the
section "Reversing Flow Direction" in the Instruc-
tion Manual of AM11 Magnetic Converter.
(2) Welding of Mating Ferrule to Piping
IMPORTANT
Weld the attached ferrules to the piping. Please
pay attention to the welding condition to avoid
deforming piping or making the stagnant portion
of the fluid; joint preparation, level difference in
butt joint, welding current.
(3) Mounting Gasket
Mount the attached gasket so it fits into the groove on
the ferrule.
(4) Positioning Flow Tube
Position the flow tube between the mating ferrule.
(5) Tightening Clamp
in-Ibf
522 to 646
Mount the clamp so they cover the flow tube and
593 to 743
mating ferrule tapered part.
T030313.EPS
Figure 3.3.8
3-15
Clamp
Welding
Mounting Procedure
3. INSTALLATION
Ferrule
Pipe
F030308.EPS
IM 1E6D0-01E