Calibration order is important to ensure the accuracy of the YK801B. The instrument needs to be
calibrated on a regular basis including after any maintenance has been performed on the probe.
The frequency at which calibration is required will depend on the specific application for which the
instrument is to be used. The optimum time between calibrations can be established by regularly
checking the performance of the instrument in standard solutions. If the YK801B is kept well
maintained and calibrated on a regular basis, a single zero point calibration is sufficient to keep the
instrument performing to specification. However, two point calibrations are required whenever a
sensor has had any maintenance.
The calibration procedures require that the probe be immersed in distilled water for zero
calibration. Make sure that you rinse both the probe and container before each calibration. The
standard solutions are available from YEO-KAL Electronics Pty Ltd or most major scientific suppliers.
Note: Zero calibration will overcome offset due to optical obscuration. This the most important
calibration point and should be done in a large black container.
To enter the calibration menu, select Calibration in main MENU and press ENTER. Select the proper
sensor and press ENTER again. Then follow the on screen instructions to complete the
calibration(s).
The previous calibrated data can be used again for the current calibration if the data is correct.

Temperature Calibration

The temperature calibration should vary very little over the lifetime of the instrument, however it is
worth checking on the accuracy of your temperature. Temperature calibration is performed at two
o
o
temperatures. The low temperature must be between 0
and 20
C and the high temperature must
o
oC
o
be between 30
and 50
. A thermometer to at least 0.1
C accuracy and resolution must be used.
To calibrate temperature, proceed as follows:
1. Select TEMPERATURE from the Calibration.
Lo Pt Raw or Hi Pt Raw will appear.
17