System - Siemens MAMMOMAT Maintenance Instruction

Table of Contents

Advertisement

2 System

SIM
Radiation protection
Check the radiation protection shield for possible damage (n°1, see figure).
Information regarding the lead equivalent value must be legible.
SIM
C-Arm – attachment and handles
Check by pulling/pushing on both handles that no parts have got loosened (n°2, see figure).
SIM
Tube cover
Check that tube cover hasn't got loosened by pulling/pushing it in different directions (n°3, see
figure).
SIM
Table retain
Check that all tables are retained when inserted and can't be removed with very light pulling force.
Inspect connectors and connector cover integrity on the table support (n°4, see figure).
SIE
Cable
System OFF
Check the condition of the power connection cable and the corrugated tubing (n°5, see figure).
Check the condition of foot pedal switches (n°6, see figure).
PMF Rotation movements
Is the rotation smooth and noiseless through the entire range?
Is the rotation brake holding the C-Arm position?
Is the mechanical goniometer functional?
Is the C-Arm rotation disabled without table?
MAMMOMAT Balance
Maintenance Instructions
SPB7-115.831.01
Page 9 of 22
Rev.01