4 Compression System
SIM
Testing the manual movement
•
System OFF
•
Move compression plate attachment Up/Down it must move free and without friction and noise
SIM
Testing compression plate support
•
Check that levers and retain pins are not damaged and release normally.
•
Check for fixation of each compression plate
SIM
Testing lexan of compression plates for wear and tear
•
Check each plate for possible damage that can result harmful to patients.
SIM
Check foot pedal switches functionality
•
Check for each pedal switch functionality
SIM
Testing magnification support fixation
•
Check that the magnification plateform can be fixed at 1,5x and 2x
SIM
Testing the "Compression" force
•
Use shifted compression plate, position a bathroom balance on the top of the table and tennis ball,
select maximum compression force and check it.
SIM
Testing the "Decompression" force
•
Move compression plate up and check that force to stop it by hand is very light.
PMF Compression Thickness Display
•
With large focus selected and compression plate in place check for compression thickness
indication using Plexiglas phantom blocks to simulate breast thickness.
•
Repeat with small focus and magnification plateform + compression plate
If thickness value is not correct refer to service instructions SPB7-115.840.01..., chapter of dosimeter
where source to skin distance calibration is described.
Breast thickness is obtained by such value subtracted by F.F.D.
MAMMOMAT Balance
Maintenance Instructions
SPB7-115.831.01
Page 11 of 22
Rev.01
05.04