4 Compression System

SIM
Testing the manual movement
System OFF
Move compression plate attachment Up/Down it must move free and without friction and noise
SIM
Testing compression plate support
Check that levers and retain pins are not damaged and release normally.
Check for fixation of each compression plate
SIM
Testing lexan of compression plates for wear and tear
Check each plate for possible damage that can result harmful to patients.
SIM
Check foot pedal switches functionality
Check for each pedal switch functionality
SIM
Testing magnification support fixation
Check that the magnification plateform can be fixed at 1,5x and 2x
SIM
Testing the "Compression" force
Use shifted compression plate, position a bathroom balance on the top of the table and tennis ball,
select maximum compression force and check it.
SIM
Testing the "Decompression" force
Move compression plate up and check that force to stop it by hand is very light.
PMF Compression Thickness Display
With large focus selected and compression plate in place check for compression thickness
indication using Plexiglas phantom blocks to simulate breast thickness.
Repeat with small focus and magnification plateform + compression plate
If thickness value is not correct refer to service instructions SPB7-115.840.01..., chapter of dosimeter
where source to skin distance calibration is described.
Breast thickness is obtained by such value subtracted by F.F.D.
MAMMOMAT Balance
Maintenance Instructions
SPB7-115.831.01
Page 11 of 22
Rev.01
05.04
4 Compression System
Siemens AG
Medical Solutions