Agilent Technologies 85133E Operating And Service Manual page 37

Nmd-2.4 mm -f- to 2.4 mm flexible test port return cables
Table of Contents

Advertisement

Table A-1 Connector Care Quick Reference
Do
Keep connectors clean
Extend sleeve or connector nut
Use plastic end-caps during storage
Do
Inspect all connectors carefully
Look for metal particles, scratches, and dents
Do
Try compressed air first
a
Use isopropyl alcohol
Clean connector threads
Do
Clean and zero the gage before use
Use the correct gage type
Use correct end of calibration block
Gage all connectors before first use
Do
Align connectors carefully
Make preliminary connection lightly
Turn only the connector nut
Use a torque wrench for final connect
a. Use isopropyl alcohol in a well-ventilated area, allowing adequate time for moist
alcohol to evaporate and fumes to disperse prior to energizing equipment.
85133E/F/H
Handling and Storage
Do Not
Touch mating-plane surfaces
Set connectors contact — end down
Visual Inspection
Do Not
Use a damaged connector — ever
Connector Cleaning
Do Not
Use any abrasives
Get liquid into plastic support beads
Gaging Connectors
Do Not
Use an out-of-spec connector
Making Connections
Do Not
Apply bending force to connection
Over tighten preliminary connection
Twist or screw any connection
Tighten past torque wrench "break" point
Connector Care Quick Reference
Principles of Microwave Connector Care
A-3