Nmd-2.4 mm -f- to 2.4 mm flexible test port return cables
Table of Contents

Advertisement

Use, Maintenance, and Care of the Cables
Cleaning Connectors
you reassemble or use it.
3. Clean the Mating Plane Surfaces
a. Apply a small amount of isopropyl alcohol to a lint-free cleaning swab.
b. Clean the center and outer conductor mating plane surfaces. Refer to
page
2-4. When cleaning a female connector, avoid snagging the swab on the center
conductor contact fingers by using short strokes.
c. Let the alcohol evaporate, then blow the connector dry with a gentle stream of clean,
low-pressure compressed air or nitrogen. Always completely dry a connector before
you reassemble or use it.
4. Reinspect
Inspect the connector again to make sure that no particles or residue are present.
3-6