3.10 Service Sub-menu

Service Setting
Error Code (Press the
SmartDial to access the Error
Code Menu)
NOTE: This is a read-only menu. Please refer any Error Codes to Fisher & Paykel Healthcare.

4. MULTI-PATIENT USE

The device does not require routine servicing or calibration. High-level disinfection cleaning is required when the device is
intended for multiple patient use and must be carried out by qualified personnel.

4.1 Deleting Sleep Data Between Patients

For accurate compliance reports, data must be cleared from the device memory as well as the InfoUSB before use by another
patient. Follow the instructions below to delete Sleep Data from the device memory:
• Turn the SmartDial to the sleep data setting
• Press to enter.
• Press again and hold for 3 seconds for the hours to start flashing.
• Press the SmartDial to delete the patient data, or turn to retain.
North American Models
• Turn the SmartDial to Sleep Data setting
• Turn to the Patient Report
• Press again and hold for 3 seconds for the hours to start flashing.
• Press the SmartDial to delete the patient data, or turn to retain.

4.2 Cleaning the Device

To maintain the optimal life and presentation of the device, regular cleaning is recommended. Refer to cleaning instructions in
the Use & Care Manual. For cleaning between patients, follow the instructions outlined below:
• Disconnect the device from the power supply
• Wipe the exterior of the device with a clean, damp (not wet) cloth and mild dishwashing detergent
• Remove the Chamber Lid, Water Chamber, Breathing Tube, Elbow and Outlet Seal
• Wipe the inside basin of the device with a clean, damp (not wet) cloth and mild dishwashing detergent
• The one-piece Water Chamber should not be disassembled.
NOTE: Do not use harsh abrasives or solvents as these may damage the device.
Remove Chamber Lid
CPAP Model Identifier
In this example the model is an Auto
with ThermoSmart™ and SensAwake
Technology
Total Non-resettable
Run Hours
while in the Clinician Menu.
while in the Clinician Menu.
or CMS report
and press to enter.
Remove Water Chamber
Remove Breathing Tube
Remove Elbow
A-7
Current Software Version
Return to the Clinician Menu
Remove Outlet Seal