Troubleshooting The Adc12Dj5200Rfevm; Table 5-1. Troubleshooting - Texas Instruments ADC DJ 00RF Series User Manual

Evaluation module
Hide thumbs Also See for ADC DJ 00RF Series:
Table of Contents

Advertisement

www.ti.com

5 Troubleshooting the ADC12DJ5200RFEVM

Table 5-1
lists some troubleshooting procedures.
Issue
General problems
TSW14J57 LEDs are not
correct
Configuration GUI is not
working properly
Configuration GUI is not able
to connect to the EVM
HSDC Pro software is not
capturing good data or
analysis results are incorrect.
HSDC Pro software gives a
time-out error when capturing
data
Sub-optimal measured
performance
SLAU640A – APRIL 2019 – REVISED JUNE 2021
Submit Document Feedback

Table 5-1. Troubleshooting

Verify the test setup shown in
document.
Check power supply to EVM and TSW14J57EVM. Verify that the power switch is in the on position.
Check signal and clock connections to EVM.
Visually check the top and bottom sides of the board to verify that nothing looks discolored or damaged.
Ensure the board-to-board FMC+ connection is secure.
Try pressing the CPU_RESET button on the TSW14J57EVM. Also try clicking Instrument Options →
Reset Board after changing the ADC configuration.
Try power-cycling the external power supply to the EVM, and reprogram the LMK and ADC devices.
Verify the settings of the configuration switches on the TSW14J57EVM.
Verify that the clock going to the CLK input is connected and the appropriate LEDs are blinking.
Verify that the ADC device internal registers are configured properly.
If LEDs are not blinking, reprogram the ADC EVM devices.
Try pressing the CPU_RESET button on the TSW14J57EVM.
Try capturing data in HSDC Pro to force an LED status update
Verify that the USB cable is plugged into the EVM and the PC.
Check the computer device manager and verify that a USB serial device is recognized when the EVM is
connected to the PC.
Verify that the green USB Status LED light in the top right corner of the GUI is lit. If it is not lit, click the
Reconnect FTDI button.
Try restarting the configuration GUI.
Use the free FT_PROG software from FTDI chip and verify that the onboard FTDI chip is programmed
with the product description ADC12DJxx00RF .
Verify that the TSW14J57EVM is properly connected to the PC with a mini-USB cable and that the
board serial number is properly identified by the HSDC software.
Check that the proper ADC device mode is selected. The mode should match in HSDC Pro and the
ADC GUI.
Check that the analysis parameters are properly configured.
Try to reprogram the LMK device and reset the JESD204 link.
Verify that the ADC sampling rate is correctly set in the HSDC software.
Try pressing the CPU_RESET button on the TSW14J57EVM. Also try clicking Instrument Options →
Reset Board after changing the ADC configuration. Try to recapture again.
Select Instrument Options → Download Firmware and download
'TSW14J57RevE_16L_XCVR_ADCDDRDACBRAM.rbf'. Try to capture again.
Try clicking Cal Triggered/Running button 2× to re-calibrate the ADC in the current operating conditions.
It is located on the Control tab of the configuration GUI.
Check that the spectral analysis parameters are properly configured.
Verify that bandpass filters are used in the clock and input signal paths and that low-noise signal
sources are used.
Copyright © 2021 Texas Instruments Incorporated
Troubleshooting the ADC12DJ5200RFEVM
Troubleshoot
Figure
3-1, and repeat the setup procedure as described in this
ADCxxDJxx00RF Evaluation Module
15