52 | DR 100e (Digital Radiography) | Introduction

Disinfecting safety directions

To disinfect the device:
• Do not use any corrosive, soluble or gaseous disinfectants.
• Use of spray disinfection can cause malfunctions due to ingress of the
disinfectant into the equipment. Disinfect all parts of the unit, including
the accessories and connection cables by just wiping them. Switch off the
system and cover the cooled system carefully before performing a room
disinfection using nebuliser.
• Using unsuitable disinfectants can cause discoloration and damage of the
surface of the equipment.
0365B EN 20181015 0853
WARNING:
Using a disinfectant that can form an explosive or flammable
gas mixtures is hazard to life and health because of explosion
risk. Switch the equipment off before disinfecting. Allow the gas
mixture to evaporate before switching the x-ray system back on.