1

Intended use

This device is intended to be used for the capture of digital images, but not for clinical diagnostic purposes.
USA: CAUTION:
Federal law restricts this device to sale by or on the order of an appropriately licensed healthcare practitioner.
2

Conformity of the System

Restrictions in Use
1. The applicable camera adapter is the U-TV0.5XC-3, U-TV0.63XC, MVX-TV0.63XC, U-TV1XC or the combination of U-
TV1X-2 and U-CMAD3.
The U-TV0.5XC should not be used because it deteriorates the image flatness.
A camera adapter with a magnification below 0.5X cannot be used because part of image will be obscured.
2. When the DP73 is connected to the rear port of the U-DPT(double port tube) or U-MPH(multi port head), the periph-
eral part of the recorded image may be deteriorated due to the optical performance of the U-DPT or U-MPH.
3. When the U-TV0.5/C-2/U-TV0.5/C-3/U-TV0.63XC (C mount camera adapter) is used, using two or more intermediate
attachments* may obscure the peripheral part of the field of view or may make flare noticeable.
* Example of two intermediate attachments with BX microscope:
Reflected light illuminator + intermediate attachment equivalent to the U-CA.