OXY5500 QF Gas Analyzer XA02754C Table of Contents Introduction ....... . . 1 How to Use This Manual.
Page 6
XA02754C OXY5500 QF Gas Analyzer Prepare, clean, decontaminate the analyzer for shipment/storage ......12 Cleaning the Instrument ..... . . 13 Maintenance and Service .
OXY5500 QF Gas Analyzer XA02754C Introduction Endress+Hauser’s OXY5500 QF gas analyzer is a stand-alone oxygen meter designed to detect oxygen in gases such as Natural Gas and air. Its design is based on fluorescence quenching technology that creates very stable, inter- nally referenced measured values.
• Refer to the Endress+Hauser website (https://endress.com/contact) for the list of local sales channels to request order-specific documen- tation. • For standard orders: • Refer to the Endress+Hauser website product page to download the published manuals for the analyzer: www.endress.com Manufacturer Endress+Hauser Address...
OXY5500 analyzer. Any use of the equipment in a manner not specified by Endress+Hauser could impair the protection provided by the equipment. Warning Labels and...
XA02754C OXY5500 QF Gas Analyzer Toxins. Endress+Hauser analyzers measure a variety of gases, including high-level H S. Follow all safety protocols governing toxic gases and potential leaks. Inhalation. Inhaling toxic gases or fumes may cause physical damage or death. Technicians are expected to follow all safety protocols established by the customer that are necessary for servicing or operating the analyzer.
Maximum voltage and current specifications for fuses. Analyzer Technical For the OXY5500 QF Gas Analyzer, a set of technical specifications are pro- Specifications vided that outline recommended equipment settings and ratings. Standard system drawings are included on → 16 that illustrate the OXY5500 ana- lyzer outline and dimensions.
Page 12
XA02754C OXY5500 QF Gas Analyzer Table 2–1 OXY5500 analyzer specifications Application Data Target Components Typical OP-9 OP-6 OP-3 Measurement 0 to 200 ppmv 0 to 5% 0 to 20% Ranges (default) 0-1 to 0- 0-10 to 0-20%...
OXY5500 QF Gas Analyzer XA02754C Sample Probe 316 Stainless Steel Construction Area CSA: Class I, Div. 2, Groups A, B, C and D T4, NEMA 4X Classification — ATEX/IECEx: II 3 G, Ex nA IIC T3 Gc IP64, Certification LCIE 16 ATEX 1012 X, IECEx LCIE 16.0009 X Technical notes •...
XA02754C OXY5500 QF Gas Analyzer Electrocution hazard Shut off power at the main disconnect external to the analyzer. Complete this action before performing any service that requires working near the main input power or disconnecting any wiring or other electrical components.
(L/R ratio) for the field wiring interface must be less than 25 μH/Ω. Mounting the The OXY5500 QF Gas Analyzer is manufactured for wall installations. Analyzer Depending on your application and configuration, the analyzer will come mounted on a plate. For detailed mounting dimensions, refer to the layout diagrams provided with the OXY5500 analyzer or in the OXY5500 Operating Instruction →...
XA02754C OXY5500 QF Gas Analyzer Opening/Closing the Analyzer Enclosure Hazardous voltage and risk of electric shock. Failure to properly ground the analyzer may create a high-voltage shock hazard. Conduit seals or Ex e cable gland should be used where appropriate in compliance with local regulations.
Page 17
OXY5500 QF Gas Analyzer XA02754C The customer-provided circuit breaker shall meet the relevant requirements of IEC 60947-1 and IEC 60947-3 and must be marked as the disconnecting device for the analyzer. This switch or breaker shall not interrupt the protective earth conductor.
This chapter provides an overview of safety operational instructions for the OXY5500 QF Gas Analyzer. Operating Controls The OXY5500 QF Gas Analyzer is shipped with complete operational instruc- tions and drawings. For operational control instructions → 1. Intermittent If the analyzer is to be stored or shut down for a short time period, follow the Operation instructions for shutting down the sample conditioning system (SCS).
OXY5500 QF Gas Analyzer XA02754C 17. Cap all inlets and outlets to prevent foreign material such as dust or water from entering the system). 18. Ensure the analyzer is free from dust, oils or any foreign material → 19. Pack the equipment in the original packaging in which it was shipped, if available.
XA02754C OXY5500 QF Gas Analyzer Safety Instruction Maintenance and Service This chapter provides safety information related to maintenance and service of the OXY5500 analyzer. Fuse Ratings and Use only the same type and rating of fuse for replacements. Characteristics Refer to the specifications listed in OXY5500 analyzer specifications →...
Page 21
OXY5500 QF Gas Analyzer Safety Instruction XA02754C Figure 5–1 Inside cabinet view (AC version) 1. AC/DC Power Connection 5. Optical Module with SMA Connector 2. RJ-45 and USB Connectors 6. Fuse Enclosure 3. Relays 7. Electro-Optical Module 4. Protective Earth Ground...