4.6.7 LED Tests

4.6.8 External DC Verification

4.6.9 NIBP Determination

Revision A
Calibration & Maintenance: 4.6 ProCare Patient Monitor Parameter Tests
1. Power on the ProCare Monitor.
2. During the power-up self-test verify all 7 segment LED display segments
and all discrete LEDs (except CHARGING LEDs) illuminate and are the
correct color.
RED
Systolic
Diastolic
°F
°C
HIGH (all)
LOW (all)
Alarm Bell Icon
SpO
Strength Meter
2
3. Repeat power up cycle until all LEDs are checked.
1. Plug the DC power cable into the monitor.
2. Verify that the CHARGING indicator is illuminated.
1. Set NIBP Analyzer to Adult Mode:
SYS/DIA = 120/80
MAP = 90
BPM = 80
2. Press START/STOP button on the Monitor to begin determination
3. Record and verify systolic, diastolic, map and heart rate from the monitor
display
4. Press CYCLE button to initiate a determination in "Auto BP" mode.
ProCare Patient Monitor Service Manual
LED COLOR MATRIX
YELLOW
Pulse Rate
LOW
2009381-001
GREEN
NOT LIT
AUTO CYCLE
CHARGING
HISTORY
BATTERY
BATTERY ICON
ALARM VOLUME
PULSE VOLUME
INFLATE
PRESSURE
4-15