Service Requirements

Disassembly/Assembly

Guidelines

Required Tools

Before Disassembly

8-2
Disassembly: Service Requirements
Follow the service requirements listed below.
Refer equipment servicing to GE authorized service personnel only.
Any unauthorized attempt to repair equipment under warranty voids
that warranty.
It is the user's responsibility to report the need for service to GE or to
one of their authorized agents.
Failure on the part of the responsible individual, hospital, or
institution using this equipment to implement a satisfactory
maintenance schedule may cause undue equipment failure and
possible health hazards.
Regular maintenance, irrespective of usage, is essential to ensure
that the equipment will always be functional when required.
WARNINGS
REPAIR TO THE FRU LEVEL — Field repairs are
recommended to the field replaceable unit (FRU) only.
Attempting a field repair on a PCB or a factory sealed
component or assembly could jeopardize the safe and
effective operation of the device.
NOTE
GE recommends that you assemble the devices using the NEW
screws with a thread lock coating provided in the hardware kit.
Screws with thread lock coating, are NOT intended to be re-used
more than four times.
NOTE
Screw part numbers are given in the FRU assembly procedures in
case of damage or loss. If a replacement screw is needed, use the part
numbers to locate the correct screw in the hardware kit.
A standard set of hand tools is required for disassembly and assembly.
Before you disassemble any devices, ALWAYS do the following tasks.
1. Power down and unplug the device from the AC power source.
2. Remove any batteries.