Attaching and adjusting the camera adapter

Attaching to the camera port on the front side of the microscope
1
a
2
3
f
4
5
a
6
TIP
Attaching to the side port unit or the trinocular tube
Insert the Allen screwdriver provided with microscope frame into the
clamping screw hole a of the camera port on the front side of the
microscope, loosen the inside clamping screw, and remove the cap b .
Screw the TV camera d into the C-mount of the camera adapter
c firmly.
Insert the C-mount part so that the parfocality adjustment screw e
matches with the adjustment hole f , fit the Allen screwdriver into the
clamping screw hole a and tighten the screw to clamp.
Insert the Allen screwdriver into the adjustment hole f and loosen
the adjustment screw e .
Set the main switches of the microscope and TV camera to (ON)
and bring the sample into focus precisely by observing it through the
eyepieces.
Rotate the TV camera d so that the image on the TV monitor is
brought into focus, and then, tighten the adjustment screw e at the
position where it is in focus.
If it is required to adjust the orientation of the TV camera,
loosen the screw inside the clamping screw hole a , adjust
the orientation and tighten the screw.
For details, refer to the instruction manuals for the camera
and the camera adapter.
For details, refer to the instruction manual provided with the control
box (BX3M-CBFM) or the coded function system (U-CBS).