• Failure to properly clean the COMP A-I-R components may cause bacterial growth, serious illness from contamination, compromised medication
delivery.
• Clean the parts of this device before using this device for the first time and also when the device has not been used for an extended period of time.
• Clean the parts of this device after each use.
• After cleaning, allow the parts to dry before reassembling. To avoid contamination, store the device and parts in a clean and safe location.
• DO NOT leave the cleaning solution on the parts. Rinse the parts with clean tap water after cleaning.
CAUTION
• Dry the parts by air drying only. DO NOT use any equipment to dry the device parts.
• To prevent damage to the device, DO NOT immerse the main unit in any liquid, DO NOT wash or rinse any of the parts under strong running water.
• To prevent damage to the device, DO NOT clean the main unit using abrasive cleaners or any type of chemical.
• NEVER clean with benzene or thinner.
Cleaning the Device Before First Use
1
Remove the inhalation top and vaporizer head from the
medication tank.
Remove any accessories (mouthpiece, mask or child accessory) from the
nebulizer kit if already attached.
2
Remove the vaporizer head from the inhalation top.
Summary of Contents for Omron COMP A-I-R RE-C800KD
Page 1
Compressor Nebulizer RE-C800KD Cleaning Instructions 5636423-9A CLEAN PARTS BEFORE FIRST USE WARNING • Failure to properly clean the COMP A-I-R components may cause bacterial growth, serious illness from contamination, compromised medication delivery. • Clean the parts of this device before using this device for the first time and also when the device has not been used for an extended period of time. •...
Page 2
Wash the parts. Wash in warm water with mild (neutral) detergent Wash the inhalation top, medication tank and accessories (mouthpiece, mask or child accessory) in warm water with mild (neutral) detergent and rinse all pieces thoroughly with clean hot tap water. Wash the vaporizer head with running tap water.