Sunrise Medical Quickie Pulse Instruction Sheet

Sunrise Medical Quickie Pulse Instruction Sheet

Rnet controller replacement instructions

Advertisement

Quick Links

2842 Business Park Ave.
Fresno, California 93727-1328
(559) 292-2171 (800) 456-8168
Note: All programming of profiles and preferences are stored within the motor controller. It is recommended that the program be "Read" or save prior to
replacement. Once the controller has been replaced the stored program can be "Written" or loaded back into the controller. This process can be accomplished
with a DTT or PC programmer.
1. Disconnect battery power prior to servicing the motor controller.
a. Carefully remove seat. (See owners manual Sec. I. Seat Removal).
b. Remove the shroud by first removing the six fasteners (Fig. 1) using a flat head screwdriver as shown (Fig. 2). Then lift up on the shroud.
c. Unplug batteries from the chair (Fig. 3).
2. The controller is mounted at the rear of the chair (Fig. 4). Firmly lift up on the controller. The controller mounts securely using 2 locating pins.
3. Unplug the 2 motor connectors, battery connector, bus connector, and inhibit/actuator connector. (Connectors have locking latches that require releasing
by squeezing with fingers while pulling connectors.)
Fig. 1
4. Remove the controller.
5. Your replacement controller will arrive with the mounting plate already mounted. Install replacement controller by reversing instructions. (Ensure that
connectors are firmly inserted until the latch "clicks" into place.
6. Once the controller is fully installed power the controller on via the remote or OMNI. (There will be an initialization "hour-glass" then the chair will be
ready to use).
FOR DEALER USE Installed by:
112675 REV.A
INSTRUCTION SHEET
Pulse Rnet Controller Replacement Instructions
Fig. 2
Fig. 3
Date:
April 2009
Connectors
Fig. 4
(retain as record of installation)
Page 1 of 2

Advertisement

loading

Summary of Contents for Sunrise Medical Quickie Pulse

  • Page 1 INSTRUCTION SHEET Pulse Rnet Controller Replacement Instructions 2842 Business Park Ave. Fresno, California 93727-1328 (559) 292-2171 (800) 456-8168 Note: All programming of profiles and preferences are stored within the motor controller. It is recommended that the program be “Read” or save prior to replacement.