UUT Range
10.99 Hz
1099.9 Hz
21.999 kHz
50 kHz
10.99 Hz
1099.9 Hz
UUT Range
109.99 Hz
22
TEKTRONIX FCA3000 Counter
Fluke 123
Use Channel A
BNC Input
Figure 9. Frequency Source Verification Connections
Table 12. Frequency Source Verification Points
Frequency @ 1 Vpp
5 Hz Sine
1 kHz Sine
10 kHz Sine
49 kHz Sine
5 Hz Square
1 kHz Square
Frequency @ 7.5 Vpp
50 Hz Square
Fluke 75x
Minimum
Frequency
4.99 Hz
5.01 Hz
999.9 Hz
1000.1 Hz
9.998 kHz
10.002 kHz
48.995 kHz
49.005 kHz
4.99 Hz
5.01 Hz
999.9 Hz
1000.1 Hz
Minimum Frequency
Maximum Frequency
49.9 Hz
50.1 Hz