Fluke 7341 User Manual page 83

Calibration bath
Hide thumbs Also See for 7341:
Table of Contents

Advertisement

Set CG calibration
parameter to n
Command
Description
co[ol]
Read Cool mode
co[ol]=of[f]/on/au
Set Cool mode:
[to]
co[ol]=of[f]
Set Cool mode to Off
co[ol]=on
Set Cool mode to On
co[ol]=au[to]
Set Cool mode to
Auto
hg[b]
Read HGb mode
hg[b]=of[f]/on/au
Set HGb mode:
[to]
hg[b]=of[f]
Set HGb mode to Off
hg[b]=on
Set HGb mode to On
hg[b]=a[uto]
Set HGb mode to
Auto
*tl[ow]
Read low set-point
limit value
*tl[ow]=n
Set low set-point
limit to n
*th[igh]
Read high set-point
limit value
*th[igh]=n
Set high set-point
limit to n
Miscellaneous (not on menus)
all
Read all extended
parameters
*ver[sion]
Read firmware
version number
*all
Read all operating
parameters
h[elp]
Read structure of all
commands
Legend:
[] Optional Command data
{} Returns either information
n Numeric data supplied by user
9 Numeric data returned to user
x Character data returned to user
Note:
When DUPLEX is set to FULL and a command is sent to READ, the command is returned
followed by a carriage return and linefeed. Then the value is returned as indicated in the
RETURNED column.
*cg=n
*cg=406.25
Table 10-1. Interface Command Summary (cont.)
Command
Command
Format
Example
co
co=of
co=on
co=au
hg
hg=of
hg=on
hg=au
*tl
*tl=–40
*th
*th=150
all
*ver
*all
h
Digital Communication Interface
Returned
Returned
Example
co:xxx
co: auto
hgb:xxx
hgb: auto
tl: 999
tl: –40
th: 999
th: 150
list of extended
parameters
ver.9999,9.99
ver.7341,1.
00
list of operating
parameters
list of commands
10
Cooling Control
Unlimited
Acceptable
Values
On, Off, Auto
On, Off, Auto
–60 to 20
–150 to 30
10-9

Advertisement

Table of Contents
loading

Table of Contents