Cleaning And Maintenance; Cleaning; Fuse Replacement - Teledyne Everywhereyoulook MX 62 User Manual

Wb interface wb/4-20ma converter
Table of Contents

Advertisement

7

Cleaning and maintenance

Cleaning

Do not use alcohol- or ammonia-based liquids to clean the controller. If necessary, clean the
exterior of the enclosure with a damp cloth.

Fuse replacement

Fuse replacement should only be performed by qualified personnel and power must
be first switched off.
Fuses shall comply with IEC 60127 standard (time-delay fuse, low breaking capacity, 250Vac).
Procedure:
• remove the rear cover from the unit by pushing aside the four clips at the corners of the
housing,
• extract the circuit from the housing,
• change the fuse shown in Figure 5/Ref. 1,
• refit all parts
Description
Fuse (1,25A- 250Vac)
NPMWBEN
Revision A.0
Figure 5
WB Interface
WB/4-20mA CONVERTER
USER MANUAL
PN
6154624
15

Advertisement

Table of Contents
loading

Table of Contents