DANGER!
Risk of Injury or Death
While Invacare strives to produce the best oxygen
concentrator in the market today, this oxygen
concentrator can fail to produce oxygen due to
power failure or device malfunction.
– ALWAYS have a backup source of oxygen readily
available.
– In the event the concentrator fails to produce
oxygen, the concentrator will briefly alarm
signaling the patient to switch to their backup
source of oxygen. Refer to Troubleshooting for
more detail.

1.3 Indications For Use

To provide supplemental oxygen to patients with respiratory
disorders, by separating nitrogen from room air, by way of a
molecular sieve.

1.4 Description

The Invacare Perfecto
V Oxygen Concentrator is used by
2
patients with respiratory disorders who require supplemental
oxygen. The device is not intended to sustain or support life.
The oxygen concentration level of the output gas ranges
from 87% to 95.6%. The oxygen is delivered to the patient
through the use of a nasal cannula.
The Invacare Perfecto
V Oxygen Concentrator uses a
2
molecular sieve and pressure swing adsorption methodology
to produce the oxygen gas output. Ambient air enters the
device, is filtered and then compressed. This compressed
1195056-B
air is then directed toward one of two nitrogen adsorbing
sieve beds.
Concentrated oxygen exits the opposite end of the active
sieve bed and is directed into an oxygen reservoir where it
is delivered to the patient.
The Invacare Perfecto
V Oxygen Concentrator is capable of
2
operation by the patient in a home environment or in an
institutional environment. Device operates at a nominal 230
VAC/50 Hertz supply.
Service information will be available upon request to
qualified technical personnel only.

1.5 Contraindications

There are no known contraindications.