Invacare® Perfecto₂™ V Oxygen Concentrator

10 Technical Data

10.1 Specifications

Electrical Requirements:
Rated Current Input:
Sound Pressure Level:
Sound Power Level:
Altitude:
Oxygen Output Concentration
Levels:
Maximum Outlet Pressure:
Flow Range:
Potential Obstruction Alert:
Power Consumption:
36
230 VAC +10, -15% (253 VAC/195.5 VAC), 50 Hz
1.6 A
< 40 dBA weighted for 3 L/min flow
< 40 dBA weighted for 5 L/min flow
< 40 dBA weighted for 5 L/min flow
Up to 1828 m (6,000 ft) above sea level without degradation of concentration levels.
Atmospheric pressure range: 101.33 kPa — 81.22 kPa
87% to 95.6% at 0.5 to 5 L/min
93% ± 3% at 2 L/min
Concentration levels achieved after initial warm-up period (approximately 30
minutes)
34.5 kPa ± 3.45 kPa (5 psi ± 0.5 psi)
0.5 to 5 L/min (maximum). For flowrates less than 1 L/min, we recommend the use of
the Invacare Pediatric Flowmeter Accessory (IRCPF16AW).
0 L/min to 0.5 L/min
The concentrator detects a condition that may indicate a potential obstruction of the
output oxygen. Rapid audible beeping alert (this alert is deactivated when accessories are
connected). May be associated with flow setting of 0.5 L/min or less.
340 W typical at 230 VAC, 50 Hz
1195056-B