You can push  to show the necessary input signal level and G to
show the necessary input signal frequency for the calibration step.
8. Turn on the output signal on the calibrator.
9. Push D to complete the step and move to the subsequent step.
If the step shown in the display moves to the subsequent step, the adjustment was
successful. If the Meter gives two beeps and does not move to the subsequent step,
then the adjustment was not successful. Make sure the calibrator is set correctly and
push D again. If the calibrator output is correct, then it is necessary to repair the
Meter.
If the calibration adjustment is not completed correctly, the Meter will not
operate correctly.
10. Put the calibrator in standby.
11. Do steps 6 through 10 for each calibration step in Table 7.
12. After you complete the last adjustment step, End shows in the display. Push D to
complete the calibration adjustment procedure and store the calibration constants in
the Meter.
Function
(Switch Position)
(AC Volts)
(DC Volts)
(DC Millivolts)
(Ohms)
 (Diode Test)
Note
Note
Table 7. Calibration Adjustment Steps
Adjustment Step
C-01
C-02
C-03
C-04
C-05
C-06
C-07
C-08
C-09
C-10
C-11
C-12
C-13
C-14
C-15
C-16
C-17
C-18
C-19
C-20
C-21
Input Value
600.0 mV, 60 Hz
600.0 mV, 20 kHz
6.000 V, 60 Hz
6.000 V, 20 kHz
60.00 V, 60 Hz
60.00 V, 20 kHz
600.0 V, 60 Hz
600.0 V, 10 kHz
6.000 V
60.00 V
600.0 V
600.0 mV
60.00 mV
600.0 Ω
6.000 kΩ
60.00 kΩ
600.0 kΩ
6.000 MΩ
0.000 Ω
50.0 MΩ
3.000 V
Digital Multimeter
Calibration Adjustments
17