GE Bently Nevada 3300/25 Operation Manual page 50

Dual accelerometer input monitor
Table of Contents

Advertisement

3300/25 Dual Accelerometer Input Monitor Operation Manual
2.
Read any other stored error codes
by pressing and holding the ALERT
switch for approximately one
second.
For example, the display to the far
right contains a second stored
error code - number 10.
When you reach the end of the
error code list, the LCD bargraph
comes on at full scale range and
the OK LEDs turn off.
You may read through the list
again by continuing to press the
ALERT switch.
3.
When the LCD bargraph is at full
scale, clear error codes from
memory by pressing and holding
the DANGER switch for
approximately one second.
If Timed OK /Channel Defeat is
selected, 30 seconds after you
clear the error codes, the OK LEDs
will flash at 1 Hz to indicate that
the monitor has been not OK.
44

Hide quick links: