HP 8562E Service Manual page 44

Spectrum analyzer
Table of Contents

Advertisement

Assembly Changed
or Repaired
A2 controller
A3 interface
A4 log amp/Cal osc
A5 IF
A6 power supply
A7 switched LO
distribution amplifiei
2-4 Adjustment Procedures
Table 2-1. Related Adjustments
Perform the following related adjustments in the order listed
16 MHz PLL adjustment
Display adjustment
new A2 or if EEROM must be replaced, also perform
the following adjustments:
LO distribution amplifier adjustment
External mixer amplitude adjustment
YIG-tuned filter/mixer adjustment
Frequency response adjustment
Display adjustment (fast zero span)
Frequency response adjustment
Display adjustment (Fast Zero Span)
Demodulator adjustment
DC log amplifier adjustment
High voltage power supply adjustment
Display adjustment
High voltage power supply adjustment
Display adjustment
LO distribution amplifier adjustment
Frequency response adjustment (or perform the
frequency response performance test in the
HP 8560 E-Series Spectrum Analyzer Calibration Guide.
The adjustment must be performed if the performance test fails.:
4djustment
Number
17
2
8
15
16
9
2
9
2
13
4
5
3
4
1
2
1
2
8
9