Available languages

Error 4
What You Should Do
1.
Conduct a control solution test using a new test strip. If the results
of the control solution test are within the range printed on the side
of your test strip vial, retest using blood and a new FreeStyle® Lite
test strip.
2.
If the retest using control solution does not work or the error
persists, call Customer Care.
Possible Causes
Problem with FreeStyle Lite test strip.