Figure 5 Alarm Setting Menu
a The high/low limit of alarm setting
In alarm setting menu,short press menu button to move the choice bar to
"Dir" item,long press menu button to choose Up or Down.Short press
menu button to adjust alarm item:SpO
low limit (SpO
ALM LO), Pulse rate high limit (PR ALM HI), Pulse rate
2
low limit (PR ALM LO).Long press menu button to change the value.Long
press menu button once,and the value raise or descend once.The low limit
can't be above high limit,and the high limit can't be below the low limit.
If the alarm function is on, the device will provide
medium-priority alarm signal when the data of SpO
beyond the limit. Intermittent alarm will occur and the values show in
yellow.
Medium priority indicating that prompt operator response is
required.
b The alarm state setting
In alarm setting menu,short press menu button to move the choice bar to
the "Alarm" item, then long press menu button to turn on or turn off the
alarm sound.Choose "on "to turn on the alarm, and choose "off" to turn off
the alarm.
c Exit operation
In alarm setting menu,short press menu button to move the choice bar to
"Exit" item, then long press button to exit alarm setting menu and return to
the main menu.
C Data Record Setting
In the main menu interface,short press menu button to move the choice bar
to the "Record" item,then long press menu button to enter the record
high limit (SpO
ALM HI), SpO
2
2
or pulse rate is
2
12