Settling Time

In normal and x2 mode, manual filter, 10 dB decreasing power step (not
across the switching point), the settling time is shown in

Table 2-7 Settling time in normal and x2 mode

Settling Time (s)
(Normal mode)
Response Time
(s) (x2 mode)
Figure 2-15 Autofilter, default resolution, 10 dB decreasing power step (not across the
N8480 Series Power Sensors Operating and Service Guide
Number of
1
Averages
0.15
0.15
0.18
Normal Mode
Maximum sensor power
150 ms
20 dB
Typical
10 dB
200 ms
Settling
Times
1 s
10 dB
6.6 s
10 dB
5 dB
13.5 s
Minimum sensor power
switching point)
2
4
8
16
0.2
0.3
0.5
1.1
0.22
0.35
0.55
Sensor
Dynamic
Range
Specifications and Characteristics
Table 2-
32
64
128
1.9
3.4
6.6
1.1
1.9
3.5
x2 Mode
Maximum sensor power
20 dB
150 ms
Typical
10 dB
180 ms
Settling
Times
400 ms
10 dB
10 dB
3.6 s
5 dB
6.6 s
Minimum sensor power
2
7.
256
512
1024
13
27
57
6.9
14.5
33
Sensor
Dynamic
Range
37