Mv Measurements - Hach sensION+ MM340 Manual

Hide thumbs Also See for sensION+ MM340:
Table of Contents

Advertisement

4. Operation
4.10. mV measurement
The instrument factory configuration is:
Channel 1, pH .
Channel 2, ISE .
Anyone of the 2 channels can measure pH, mV or ISE .
The measurement units are selected in System .
Measurement conditions
Resolution: 1 mV / 0 .1 mV
Measuring mode: By stability, In continuous or By time.
For more details, see "pH measurement", page 10 .
10-04-2010
17:15
mV MEASURE
Channel 1
...by stability
...
Start Measurement
Channel 1
Measurement in continuous
MEASURING
MEASURING
by stability
* * * * mV
510 mV
Sample 1
00:01
Sample 1
etc.
mV measurement with printer
See pH measurement with printer, page 12 .
Important note:
Measuring mV the temperature is not shown . The temperature is
displayed only if a temperature probe is connected .
mV calibration
10-04-2010
17:15
Factory Adjust
...
mV CALIBRATION
Start Calibration
Channel 1
Channel 1
Factory Adjust
Compensation: 0 mV
Continue...
"Factory adjustment"
(compensation: 0 mV),
eliminates the com-
pensation in memory.
14
Reprogramming mV measurement conditions
Measuring Conditions
Resolution: 1
Meas.: stability
Start Measurement
Measuring Conditions
in continuous
Resolution: 1
Meas.: stability
00:01
Start Measurement
etc.
Measuring Conditions
Resolution: 1
Meas.: stability
Start Measurement
Modifying standard calibration
Calibration mV
Standard 220 mV
To A X Value
Data Introduciton
Factory Adjust
Calibration mV
Standard 220 mV
To A X Value
Data Introduciton
Factory Adjust
etc.
Calibration mV
Standard 220 mV
To A X Value
Data Introduciton
Factory Adjust
10-04-2010
17:15
SYSTEM
Measurement units
SYSTEM
Sample ID
Data output
Such amended measures, it is necessary to save the changes .
SYSTEM
Measurement units
Sample ID
Data output
SYSTEM
Measurement units
Sample ID
Data output
SYSTEM
Sample ID
Measuring Conditions
Data output
Resolution mV:
Type of report
1
0.1
SYSTEM
Data output
Measuring Conditions
Type of report
Measurement:
Users
By Stability
etc.
In Continuous
By time
SYSTEM
Type of report
Users
Header
SYSTEM
Users
Header
Identify sensor
SYSTEM
Header
Identify sensor
Stirrer 2
SYSTEM
Identify sensor
CALIBRATING mV
Stirrer 2
Date / Time
etc.
Use standard 220 mV
Continue...
SYSTEM
Stirrer 2
CALIBRATING mV
Date / Time
Language
etc.
Use specific standard
Continue...
SYSTEM
Date / Time
Introducing Data
Language
Display contrast
Compensation: 0
mV
Manual introduction of calibration data
for a sensor previously calibrated
SYSTEM
Language
Display contrast
Readjust temp.
SYSTEM
Display contrast
Readjust temp.
Stirring
SYSTEM
SYSTEM
Measurement units:
Measurement units:
Channel 1: pH
Channel 2: ISE
Channel 1:
Save changes
SYSTEM
Codification:
Automatic
Manual
SYSTEM
Data output
Deactivated
For printer
For PC
SYSTEM
SYSTEM
Type of report
Samples by report:
Reduced
Several
Standard
GLP
One sample
SYSTEM
SYSTEM
Users:
User 1
User 2
Yes
No
User 3
SYSTEM
Header
First/second line
_
SYSTEM
SYSTEM
Identify sensor:
Channel 1, pH
Electrode data
Channel 1, pH
Code:_
Serial no.:_
Channel 2, ISE
Save data
SYSTEM
Stirrer 2
Yes
No
SYSTEM
Data / Time
04-06-2010
09:30
Monday
SYSTEM
ENGLISH
GERMAN
SPANISH
FRENCH
ITALIAN
SYSTEM
SYSTEM
Display contrast
Display contrast
∆∆∆∆∆∆∆∆∆∆∆∆
∆∆∆∆∆∆∆∆∆∆∆∆∆∆∆
SYSTEM
Channel 1 Correction
25.0°C: 0.0 °C
85.0°C: 0.0 °C
SYSTEM
SYSTEM
Stirring
Stir
Stirrer N. 1 OFF
Stirrer N. 1 ON
25%
ON
pH
25%
OFF