Reassembling The Acquisition Module - GE CAM-14 Service Manual

Acquisition modules
Table of Contents

Advertisement

Reassembling the Acquisition Module

Strict antistatic precautions should be followed during disassembly/assembly of the
acquisition module.
1.
Apply silicone sealant to the edge (1) of the top and bottom covers.
2.
Place the main PCB in the bottom cover.
3.
Connect the host cable interface flex (2) at the main PCB by gently pressing
the connector in place.
4.
Connect the switch flex cable (3) to the host cable interface flex at the main PCB
(4) by gently pressing the connector in place.
5.
While holding the top (5) and bottom covers together, replace the four Torx
screws (6) using a #6 Torx driver. DO NOT over tighten.
6.
Perform a Functional Checkout procedure.
2049751-001C
CAM-14
/CAM-HD
®
®
Troubleshooting