Cautions - GE CAM-14 Service Manual

Acquisition modules
Table of Contents

Advertisement

3.
To disinfect, wipe with a soft, lint-free cloth moistened with an appropriate
disinfectant.
Use the following solutions, as recommended in the APIC Guidelines for Selection
and Use of Disinfectants (1996):
• Sodium hypochlorite (5.2% household bleach) minimum 1:500 dilution
• Any sodium hypochlorite wipe product that meets the previous guidelines
4.
Allow the cleaning solution/disinfectant to remain on the device for a minimum
of one minute, or per hospital guidelines.
5.
Wipe off the cleaning solution/disinfectant with a clean, moistened cloth.
6.
Dry thoroughly with a dry, lint-free cloth and let air dry for a minimum of 30
minutes before use.
DO NOT use excessive drying techniques, such as oven, forced heat, or sun
drying.
NOTE:

Cautions

• Follow the cleaning instructions exactly.
• Wring excess disinfectant from wipe before using.
• Never immerse the device, cables, or leadwires in any liquid, as this may corrode
metal contacts and affect signal quality.
• Do not allow fluid to pool around connection pins. If this happens, blot dry with a
soft, lint-free cloth.
• Never use conductive solutions or solutions that contain chlorides, wax, or wax
compounds to clean the device, cables, or leadwires.
• Never use solutions or products that contain any type of Ammonium Chloride such
as, but not limited to:
• Dimethyl Benzyl Ammonium Chloride
• Quaternary Ammonium Chloride solutions
• Abrasive cleaners or solvents of any kind
• Acetone
• Ketone
• Betadine
• Alcohol-based cleaning agents
• Sodium salts
• Never autoclave or steam clean the device, cables, or leadwires.
• Do not use until thoroughly dry.
2049751-001C
(minimum 100 ppm free chlorine) and maximum 1:10 dilution.
can be used.
Drying times may vary based on the environmental conditions.
CAM-14
/CAM-HD
®
®
Maintenance