Operating Specifications; Table 5 Receiver (Preamplifier) Specifications; Table 6 Pulser Specifications - Olympus TRPP 5810 User Manual

Pulser/preamplifier for tofd inspection
Table of Contents

Advertisement

DMTA-20070-01EN, Rev. A, October 2014
Table 4 Operating environment specifications (continued)
Wet location
Altitude
Outdoor use
Pollution level
IP rating
4.2

Operating Specifications

Table 5 on page 26 and Table 6 on page 26 detail the operating specifications of the
TRPP 5810 unit's pulser and receiver.
Voltage gain
Bandwidth (−3 dB)
Input-referred noise
Input/output HV protection
Input impedance
Output impedance
Output voltage
Crosstalk between channels
Output pulse voltage
Output impedance
Input triggering voltage
Input impedance
Probe damping
26
Chapter 4
Parameter

Table 5 Receiver (preamplifier) specifications

Parameter
40 dB ±0.5 dB into 50 W at 10 MHz
0.5 MHz to 30 MHz with ±10 % tolerance
−300 V peak pulsed for pulse width 500 ns
550
50
4 Vpp nominal, 3.3 Vpp min. (at 50

Table 6 Pulser specifications

Parameter
−200 V ±20 V into 50 W
−15 V to −200 V, protected up to −300 V
50
Active, 50
Value
Yes
Up to 2000 m
Yes
1
Waterproof (designed to meet IP66)
Value
nV/ Hz
2
with 50
source impedance
±55
±5
100 dB with 10 MHz sine wave
Value
6
±5
±5
)

Advertisement

Table of Contents
loading

Table of Contents