5
Setup / service program
5.1
Service mode
The Service menus CALIBRATION, DIAGNOSTICS, MISCELLANEOUS, and SETUP MENU can only
be called in service mode.
5.1.1
Service mode overview
CALIBRATION
LM
DIAGNOSTICS
LM
MISCELLANEOUS
LM
SETUP MENU
LM
Fresenius Medical Care 4008 S (Version V10) SM-EN 7/01.13
To enter the service mode, proceed as
follows:
1. The device must be switched off.
2. Set the Service switch to ON (up).
3. Turn the device on.
Warning
Setting the service switch to ON while the treatment or the cleaning
program is in progress will turn off the heater relay.
Chapter 5: Setup / service program
5-1