7.1.10 Blood leak test

Error message
F01 Blood Leak
F02 Blood Leak
F03 Blood Leak
F95 Blood Leak

7.1.11 Temperature test

Error message
F01 Temperature
F02 Temperature
F03 Temperature
Fresenius Medical Care 4008 S (Version V10) SM-EN 7/01.13
Description
Blood leak channel and dimness not in alarm-free condition during the T1 test.
– Dimness channel contaminated (calcium precipitate, etc.)
– Acknowledgement (BLL, X637/A18)  X633L/B10 voltage value within the
alarm tolerances ( 3 V /  8 V)
– Acknowledgement (BLL_DIM, X637A/21)  X633L/B11 voltage value within
the alarm tolerances ( 1.5 V /  8 V)
– DAC_BLL or DAC_DIM not within the tolerances (check calibration)
The blood leak alarm / dimness alarm is not recognized during test detuning.
– Acknowledgement (BLL, X637/A18)  X633L/B10 voltage value not within
the alarm tolerances ( 3 V)
– Acknowledgement (BLL_DIM, X637A/21)  X633L/B11 voltage value not
within the alarm tolerances ( 1.5 V)
– Test detuning (BLL_DET, X632/A25)  X633L/B27 not 5 V
– Calibration of DAC_BLL or DAC_DIM is too high
– Detuning (DAC_DIM, X634R/A11)  X633L/C3 impossible
– Dimness calibration is set to potentiometer calibration
(BR6 from pos. 1/2 to 2/3)
After test detuning, the blood leak channel and dimness fail to enter the
alarm-free state.
– Dimness channel contaminated (calcium precipitate, etc.)
– Acknowledgement (BLL, X637/A18)  X633L/B10 voltage value within the
alarm tolerances (3 V /  8 V)
– Test detuning (BLL_DET, X632/A25)  X633L/B27 not 0 V
– Acknowledgement (BLL_DIM, X637A/21)  X633L/B11 voltage value within
the alarm tolerances ( 1.5 V /  8 V)
– DAC_BLL or DAC_DIM not within the tolerances (check calibration)
System error
Description
The temperature measuring range is not set to hemodialysis.
– Control line (HOTRINSE, X634R/C24)  X639/A20 not 0 V
– Acknowledgement (HOTRINSE, X634R/C24)  X632/A26 not 0 V
The actual temperature is less than 35.0 °C (test running time > 15 minutes).
– Calibrate the temperature
– The heater rod has failed
– Acknowledgement (T_DIAL1, X633L/B16)  X632/A24, voltage got stuck.
The actual temperature is higher than 39.0 °C (test running time 15> 15
minutes).
– Calibrate the temperature
– The regulating sensor (NTC-2) is defective
– Acknowledgement (T_DIAL1, X633L/B16)  X632/A24, voltage got stuck.
Chapter 7: Error messages
7-9