Hach TitraLab Total Hardness Basic User Manual page 15

Table of Contents

Advertisement

Available languages

Available languages

2.
Place the titration cell over the magnetic stirrer.
3.
Lower the electrode into the titration cell.
Note: Make sure that the electrode cannot be damaged by the
stirring bar or come into contact with the bottom of the titration cell.
Make sure that the sensitive part of the ISE electrode is covered by
sample.
4.
Select ANALYZE TOTAL HARDNESS. "INSERT SAMPLE: 1" is
shown on the display.
5.
Select START ANALYSIS.
The following steps are carried out automatically by TitraLab:
Hardness Buffer is dispensed into the titration cell for 7
seconds
The solution is stirred for 5 seconds
Titration is then started
6.
The result is shown on the display and, depending on the setting,
sent to a printer or computer.
7.
Press OK to confirm the result.
8.
To begin another analysis, select OK. "NEXT SAMPLE" is shown
on the display.
Manual component activation
Burette
1.
Select MANUAL ACTIVATION > BURETTE and press OK to
confirm.
2.
Select FILL with the DOWN navigation key to fill the burette.
3.
Select EMPTY with the UP navigation key to empty the burette into
the titration cell.
Note: Repeat this procedure (FILL/EMPTY) multiple times to make sure
that the dispensing process functions correctly.
4.
Select RETURN with the LEFT navigation key to empty the burette
back into the titrant bottle.
Pump
1.
Select MANUAL ACTIVATION > PUMPS and press OK to confirm.
English
13
2.
Press OK to switch the pump on or off.
Note: Repeat this procedure multiple times to make sure that the
dispensing process functions correctly.
Stirrer
1.
Select MANUAL ACTIVATION > STIRRER and press OK to
confirm.
2.
Press OK to switch the stirrer on or off.
3.
Use the navigation keys UP and DOWN to select the stirring speed.

Advertisement

Table of Contents
loading

Table of Contents