Calibration and adjustments
No calibrations or adjustments are needed after this part replacement.
8-6-10-7
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
1.) Verify that all screws removed earlier have been installed.
2.) If finished, connect cables and Probes removed earlier.
3.) Power up the system to verify that it operates as intended.
4.) If a Gel Warmer is installed in the Options Holder, see: additional Verification and Functional
Checks for Gel Warmer (see:
8-6-10-8
Functional Checks
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
If all are successful, include the following debrief script: LOGIQ E9 Service Manual, Direction
5573152-100, Rev. 2. Equipment passed all required checks and is ready for use.
See:
Section
4-2-2
4-2-3
4-3-13
4-3-5
4-3-7
4-3-10
4-3-17
4-3-17-2
. 2
EV
8-6-11 "Gel Warmer replacement" on page
Table 8-115 Options Holder / Left or Right Support replacement Functional Checks
Functional Check
Power ON/Boot Up
Power shut down
Probe/Connectors Checks
B-Mode Checks
System CFM and PWD Checks
Basic Measurements
Mechanical Function Checks
Brakes and Direction Lock Checks
Chapter 8 Replacement procedures
Debrief Script
LOGIQ E9 Service Manual, Direction
5573152-100, Rev. 2. Equipment passed all
required checks and is ready for use.
LOGIQ E9 S
M
ERVICE
ANUAL
8-176).
8 - 175