Safety Instructions

In this instruction sheet, a Warning identifies conditions and
actions that pose hazard(s) to the user. A Caution identifies
conditions and actions that may damage the calibrator or the test
instruments.
To avoid electric shock or personal injury:
If the Probes are used in a manner not
specified in these operating instructions,
protection provided by the Probes may be
impaired.
Use the Probes only if qualified.
Use caution during installation and use of the
Probes; high voltages and currents may be
present in circuit under test.
Have maintenance performed by only qualified
service personnel.
Protect the Probes against water and
humidity.
Wear protective clothing and gloves as
required.
Do not install this product on live conductors.
Always de-energize the circuit under test
before installing the flexible measuring head.
Always inspect the electronics unit,
connecting cable, and flexible measuring head
for damage before using the Probe.
Do not use the Probe if damaged.
Always connect Probe to display device
before installing the flexible measuring head.
Never change batteries while measurement
head is installed on conductor.
Never connect or disconnect the external
power supply while the measurement head is
installed on a conductor.
Use only the provided original or specified
accessories.
Adhere to local and national safety codes.
Individual protective equipment must be used
to prevent shock and arc blast injury where
hazardous live conductors are exposed.
CAT III equipment is designed to protect
against the transients in the equipment in
fixed equipment installations, such as
distribution panels, feeders and short branch
circuits, and the lighting systems in large
buildings.