Weight Without Peripherals; Electrical Specifications; Power Supply; Facility Power Socket - GE LOGIQ 180 Basic Service Manual

Table of Contents

Advertisement

GE M
S
EDICAL
YSTEMS
D
2317229, R
IRECTION
EVISION
3-6-1-3

Weight without Peripherals

The Weight of LOGIQ™ 180 without peripherals
Table 3-2
3-6-2

Electrical Specifications

Electrical Specifications for LOGIQ™ 180 .
Table 3-3
GEMS P/N
H41542LA
H41542LB
H41542LC
H41542LD
3-6-2-1

Power Supply

The Following Power Line requirement should be monitored one week before Installation. We
recommend that a Drantez model 605-3 Power line Monitor with options 101 to be used.
Voltage Range
Line Frequency
Power Transients
Decaying Oscillation
Table 3-4
3-6-2-2

Facility Power Socket

A separate power outlet with a 5 amp circuit breaker for 220-240 VAC is recommended. The specific
power receptacle used depends on the country's power line standards.
This receptacle should have International Electro technical Commission (IEC) approval, or equivalent
3 - 20
3
Weight of LOGIQ™ 180
Model
LOGIQ™ 180
Electrical Specifications for LOGIQ™ 180
Voltage
220-240 VAC
220-240 VAC
220-240 VAC
220-240 VAC
Parameter
China, Europe, India, Korea
Power
China, Europe, India, Korea
Less than 25% of nominal peak voltage for less than 1 millisecond for any type of transient,
including line frequency, synchronous, asynchronous or periodic transients
Less than 15% of Peak Voltage for less than 1 millisecond
Power supply
Section 3-6 - System Configuration
Weight [kg]
72.2
Tolerances
±10%
±10%
±10%
±10%
Limits
All Applications
LOGIQ™ 180 B
S
ASIC
ERVICE
Weight [lbs]
163.17
Current
1.9-2.0 A Max.
1.9-2.0 A Max.
1.9-2.0 A Max.
1.9-2.0 A Max.
±
200-240VAC
10%
MAX 450VA
±
50/60Hz
2Hz
M
ANUAL
Frequency
50 Hz
50 Hz
50 Hz
60 Hz

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

H41542laH41542lbH41542lcH41542ld