2-2-4. Printing cannot be performed normally from video signal (NTSC/PAL)
(UP-X898MD Only)
Printing cannot be performed
normally from video signal
(NTSC/PAL).
Is the printing performed normally
from PC?
Is the connection of the video
cable between this unit and
VIDEO signal output device
normal?
Is there any noise mixed in the
printed image?
UP-D898MD/X898MD
Yes
No
Yes
Replace the MA-195 board.
(Refer to Section 3-1.)
No
Perform the flow chart of
Section 2-2-3.
Check the connection of the video
Yes
cable. If the symptom persists,
replace the video cable.
Replace the video cable with the
No
shorter one, or with the one having
less attenuation.
2-7 (E)