2.2 Packaging for Storage or Transportation

WARNING
wheels, but if it is to be lifted, it is a two-person lift.

2.2.1 Environment

The optimum conditions for ADTS 405 Mk2 equipment during transit, preparation for
calibration and repair and when being storage are detailed as below. Items in storage are
defined as being non-operational. If the ADTS becomes exposed to moisture or very high
humidity, dry as soon as possible and temporarily store in a low humidity area. The following
conditions apply for storage or shipping:
Parameter
Temperature
Altitude

2.2.2 Procedure

1.
The unit should be at zero/ambient pressure.
2.
Disconnect all hose assemblies and stow them in the shoulder bag.
3.
Switch power OFF and disconnect from the electrical power supply.
4.
Disconnect the power supply cable and the hand terminal cable.
5.
Disconnect the hand terminal cable from the hand terminal.
When using packing materials other than the original, proceed as follows:
1.
Wrap unit in polyethylene sheeting.
2.
Select a double-wall cardboard container. Inside dimensions must be at least 15 cm
greater than the equipment. The carton must meet test strength requirements of
> 125 kg.
3.
Protect all sides with shock-absorbing material to prevent equipment movement
within the container.
4.
Seal carton with approved sealing tape.
5.
Mark carton "FRAGILE" on all sides, top, and bottom of shipping container.
Note: For ADTS 405R Mk2 rack-mount units, loose or peripheral items should be placed
in a sealed polythene bag. For ADTS 405F Mk2, stow the power supply cable, hand
terminal cable and the hand terminal in the ADTS 405F Mk2 lid. Fit the lid to the unit.
6 | ADTS 405 R/F Mk2 Instruction Manual-English
The flight-line unit can be moved by use of the handle and
Table 2-1: Storage Conditions
Permitted Range
-51°C to 70°C (-60° to +158°F)
Up to 50,000 feet (15,240 metres)
© 2017 General Electric Company – All Rights Reserved.