Calibration Manual
6. Set up the function generator as follows:
7. Set the oscilloscope for 0.5 V/DIV @ 0.5 µs/DIV.
8. Trigger the waveform on channel 2.
9. Adjust the amplitude of the function generator to produce a 3 VP-P triangle wave.
10. Set the oscilloscope for 1.0 V/DIV @ 5.0 ms/DIV.
11. Trigger the waveform on Channel 1.
12. Check that the peak voltage is greater than 5.7 V and decays to less than 1.0 V
13. Adjust the function generator output so the peak voltage is 6 V.
14. Set the scope for 1.0 ms/DIV and trigger waveform.
15. Check the amplitude after 3.0 ms from the waveform peak, the voltage amplitude is
16. Re-adjust the function generator output to obtain the value recorded in step 12.
17. Set the scope for 5.0 ms/DIV.
18. Turn the inductive pickup so that "SPARK PLUG SIDE" points along the wire
20
Frequency
Amplitude
Burst Mode
Burst Count
Burst Rate
Burst Phase
See Figure 6.
between pulses. See Figure 6. Record the peak value for later use.
2.4 V +0.5/-0.8 (1.6 V to 2.9 V). (See Figure 3-5(C).)
connected to the LOW output of the function generator. Check that the waveform is
less than 2 V.
900 kHz
8.3 V
yes
1
100 Hz
-90 Degrees