UTD2000M User Manual
Appendix A: Technical Indexes
Unless otherwise stated, all technical indexes are applicable to probes with attenuation switch set to be 10×
and UTD2102CM series digital storage oscilloscopes. Those digital storage oscilloscopes must first comply with
the following two conditions in order to meet those specification standards.
■The unit must be running under a specific operating temperature for more than 30 minutes.
■If the change in operating temperature reaches or exceeds 5℃, system function menu has to be opened to
execute "auto calibration" program. Other than those marked with the word "typical", the effectiveness of all
specifications are guaranteed.
Technical Indexes
Sampling Mode
Sampling
Real‐time Sampling
Mode
Sampling
Single Channel: 1GS/s; Dual Channel (both) :
Rate
500MS/s
Average
When all channels reach N times of sampling simultaneously, wherein "N" can be any number
Value
among 2, 4, 8, 16, 32, 64, 128, 256 and 512.
Input
Input Coupling
Input Impedance
Probe Type
Probe
Bandwidth
Probe
Attenuation
100MHz
Coefficient
Bandwidth
Setting
100MHz
Maximum Input Voltage
Time Delay Between Channels (Typical)
Chapter 6 Appendix
DC, AC, GND
Bandwidth≤100MHz 1±2%ΜΩ,Parallel connection with 24±3pF
Bandwidth>100MHz 1±2%ΜΩ,Parallel connection with 18±3pF
Voltage, Current (Bandwidth ≤100MHz)
Voltage (Bandwidth>100MHz)
≤
1×, 10×, 100×, 1000× or user‐defined
>
1×、10×、100×、1000×
400V(DC+AC Peak Value, Input Impedance 1MΩ)
50ps
83
Equivalent Sampling
50GS/S