Symptomatic Failure Types; Check Unratioed Power Levels; When Using The Optional Hp 8517B Test Set - HP 85106D Manual

Network analyzer system
Table of Contents

Advertisement

Check Unratioed
Power Levels
When Using the
Optional HP 8517B
Test Set
NOTE

Symptomatic Failure Types

Refer to the following heading listed that most closely represents the
instrument problem (obvious or not). They are suggested courses of action
specific to the failure symptom.
Millimeter-Wave, Phase Lock, Power Loss, or Frequency Related
Problems.
Refer to the test titled "Check Unratioed Power Levels" on page 4-19.
Microwave, Phase Lock, Power Loss, or Frequency Related Problems

When Using the Optional HP 8517B Test Set.

To help locate the faulty instrument, the service adapter (HP part number
85102-60210) is used to emulate the operation of the source and test set and
check unratioed power levels into the HP 85102 IF/Detector.
When checking any unratioed power levels, turn the HP 8510 averaging
OFF. Refer to "Unratioed Power Test" in the HP 8517B Operating and
Service Manual. This procedure allows you to check individually the output
power level of each test set sampler/mixer assembly and its associated IF
amplifier. Depending upon the test outcome, the procedure directs you to the
most likely cause of failure.
Service and Troubleshooting
Symptomatic Failure Types
HP 85106D mm-Wave System 4-11

Advertisement

Table of Contents
loading

Table of Contents