Maximum Voltage to Earth Ground . . . . . . . . . . . . . . .1000 V
Battery Type. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .AA alkaline IEC LR6 (x4)
Battery Life. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .>15 hours in Ground Resistance measurement mode, with backlight turned off, and
Frequency Range
Filter OFF . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .40 Hz to 1 kHz
Filter ON . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .40 Hz to 70 Hz
Ingress Protection. . . . . . . . . . . . . . . . . . . . . . . . . . . . .IEC 60529: IP30 with jaw closed
LCD
Digital Reading . . . . . . . . . . . . . . . . . . . . . . . . . . . .9999 counts
Refresh Rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4/second
Temperature
Operating . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .-10 °C to +50 °C
Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .-20 °C to +60 °C
Operating Humidity . . . . . . . . . . . . . . . . . . . . . . . . . . . .Non condensing (<10 °C)
Altitude
Operating . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2000 m
Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12 000 m
Reference Temperature . . . . . . . . . . . . . . . . . . . . . . . .23 °C ±5 °C (73 °F ±9 °F)
Temperature Coefficient . . . . . . . . . . . . . . . . . . . . . . . .0.15 % x reading/ °C (<18 °C or >28 °C)
Overload Indication. . . . . . . . . . . . . . . . . . . . . . . . . . . .OL
Data Logging Capacity . . . . . . . . . . . . . . . . . . . . . . . . .>2000 data points
Data Logging Interval . . . . . . . . . . . . . . . . . . . . . . . . . .1 second to 59 minutes and 59 seconds
Safety
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .IEC 61010-1: Pollution Degree 2
Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . .IEC 61010-2-032: CAT IV 600 V / CAT III 1000 V
Current Clamp for Leakage Current
Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . .IEC 61557-13: Class 2,
Resistance to Earth. . . . . . . . . . . . . . . . . . . . . . . . .IEC 61557-5
Effectiveness of the Protective Measures . . . . . . . .IEC 61557-16: cutoff frequency 20 kHz (-3 dB)
Electromagnetic Compatibility (EMC)
International . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .IEC 61326-1: Portable Electromagnetic Environment
Group 1: Equipment has intentionally generated and/or uses conductively-coupled radio frequency energy that is necessary for the
internal function of the equipment itself.
Class B: Equipment is suitable for use in domestic establishments and in establishments directly connected to a low voltage power
supply network which supplies buildings used for domestic purposes.
Emissions that exceed the levels required by CISPR 11 can occur when the equipment is connected to a test object.
Korea (KCC) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Class A equipment (Industrial Broadcast & Communications Equipment)
Class A: Equipment meets requirements for industrial electromagnetic wave equipment and the seller or user should take notice of it.
This equipment is intended for use in business environments and not to be used in homes.
USA (FCC) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .47 CFR 15 subpart B. This product is considered an exempt device per clause
Wireless Radio
Frequency Range . . . . . . . . . . . . . . . . . . . . . . . . . .2412 MHz to 2462 MHz
Output Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . .<10 mW
Radio Frequency Certification. . . . . . . . . . . . . . . . .FCC ID:T68-FBLE IC:6627A-FBLE
4
RF mode turned off
90 % RH (at 10 °C to 30 °C)
75 % RH (at 30 °C to 40 °C)
45 % RH (at 40 °C to 50 °C)
(noncondensing)
IEC 61557-1
30 A/m
CISPR 11: Group 1, Class B, IEC 61326-2-2
15.103.