Fluke 1630-2 Calibration Manual page 15

Earth ground clamp
Hide thumbs Also See for 1630-2:
Table of Contents

Advertisement

2.
At the next calibration step, if the screen shows "mA" and "AC", apply an AC current to jaw with a 2-turn
current loop. See Figure 1. The calibrator output value should be half of the value shown on the screen. For
example, the calibrator output should be 0.100 mA for the calibration point shown on the screen below. The
current frequency should be 60 Hz.
Figure 1. Calibration with 2-Turn Current Loop
Earth Ground Clamp
Calibration Procedure
5522A Calibrator
5522A
CALIBRATOR
STBY
/ +
OPR
7
8
4
5
1
2
0
11