Specifications
Measurement Category
Operating Voltage
Operating Frequency
Hazardous Voltage
Detection
Signal Indications
Response Time
Current Output of Signal
(typical)
Signal Voltage Output
(nominal)
Range Detection (open air)
Display Size
Display Dimensions (W x H)
Display Resolution
Display Type
Display Color
Booting Time
Backlight
Operating Temperature
Operating Humidity
Storage Temperature and
Humidity
Operating Altitude
Transient Protection
Pollution Degree
IP Rating
Drop Test
Power Supply
Power Consumption
(typical)
Battery Life (typical)
Low Battery Indication
Fuse
Maximum Conductor Size
Dimensions (L x W x H)
Weight (batteries installed)
Certifications
AT-8000-R Receiver
CAT IV 600 V
0 to 600 V AC/DC
Energized: 6.25 kHz
De-Energized: 32.768 kHz
See NCV detection
Numeric bar graph display and audible beep
Smart mode: 750 mSec
Tip Sensor Energized: 300 mSec
Tip Sensor De-Energized: 750 mSec
NCV: 500 mSec
Battery monitoring: 5 Sec
N/A
N/A
Smart mode
Pinpointing: Around 1.97-in (5 cm) radius (±2%)
Direction indication: Up to 5FT (152.4cm) (±2%)
TIP Sensor: Energized
Pinpointing: Around 1.97-in (5 cm) (±1%)
Detection: Up to 22-FT (670.56cm) (±1%)
TIP Sensor: De-Energized
Detection: Up to 14-FT (426.72cm) (±5%)
NCV (40-400 Hz)
Pinpointing: Around 1.97-in (5cm) radius (±5%)
Detection: Up to 4-FT (121.92cm) (±5%)
3.5 in (89 mm)
2.76 x 2.07 in (70 x 52 mm)
320 x 240
Color TFT LCD
Yes
30 sec
Yes
-4 °F to 122 °F (-20 °C to 50 °C)
45%: -4 °F to <50 °F (-20 °C to <10 °C)
95%: 50 °F to <86 °F (10 °C to <30 °C)
75%: 86 °F to <104 °F (30 °C to <40 °C)
45%: 104 °F to 122 °F (40 °C to 50 °C)
-4 °F to 158 °F (-20 °C to 70 °C),
<95% RH
N/A
IP 52
4 x AA (alkaline or NiMH rechargeable)
4 x AA battery: 2W
Approx. 9 h
Yes
N/A
N/A
Approx. 10.92 x 4.43 x 2.55 in
(278 x 113 x 65 mm)
Approx. 1.20 lb (0.544 kg)
©2020 Amprobe
Advanced Wire Tracer - Quick Guide
AT-8000-T Transmitter
CAT IV 600 V
0 to 600 V AC/DC
Energized/Loop: 6.25 kHz
De-Energized: 32.768 kHz
> 30 V AC/DC
LEDs and audible beep
Line/phase voltage monitoring: 1 sec
Battery voltage monitoring: 5 sec
Energized circuit:
HI mode: 60 mA RMS
LO mode: 30 mA RMS
De-energized circuit:
HI mode: 130 mA RMS
LO mode: 40 mA RMS
Loop mode: 160 mA RMS
De-energized circuit:
LOW: 29 V RMS, 120 Vp-p
HIGH: 33 V RMS, 140 Vp-p
Loop model: 31 V RMS, 120 Vp-p
N/A
LEDs
N/A
N/A
LEDs
Operating mode LEDs: red
Battery status LEDs: green, yellow, red
< 2 sec
N/A
-4 °F to 122 °F (-20 °C to 50 °C)
45%: -4 °F to <50 °F (-20 °C to <10 °C)
95%: 50 °F to <86 °F (10 °C to <30 °C)
75%: 86 °F to <104 °F (30 °C to <40 °C)
45%: 104 °F to 122 °F (40 °C to 50 °C)
-4 °F to 158 °F (-20 °C to 70 °C),
<95% RH
0 to 6561 ft (2000 m)
8.00 kV (1.2/50µS surge)
2
3.28 ft (1 m)
8 x AA (alkaline or NiMH rechargeable)
Hi/Lo mode: 70 mA
Loop mode with Clamp: 90 mA
Consumption without signal
transmission: 10 mA
Hi/Lo mode: approx. 25 h
Loop mode: approx. 18 h
Yes
1.6 A, 700 V, fast-acting, Ø 6x32mm
N/A
Approx. 7.2 x 3.66 x 1.97 in
(183 x 93 x 50 mm)
Approx. 1.25 lb (0.57 kg)
For detailed specifications and ordering go to amprobe.com 6012404 A
®
CT-400 Signal Clamp
CAT IV 600 V,
CAT III 1000 V
0 to 1000 V AC
Loop Mode: 6.25 kHz
High / Low Mode: 32.768 kHz
AC current measurement: 45 Hz to 400 Hz
N/A
N/A
Instantaneous
1 mA/A for AC current measurement
with multimeter
De-energized circuit:
2.4 V RMS, 24 Vp-p
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
32 °F to 122 °F (0 °C to 50 °C)
95%: 50 °F to <86 °F (10 °C to <30 °C)
75%: 86 °F to <104 °F (30 °C to <40 °C)
45%: 104 °F to <122 °F (40 °C to <50 °C)
-4 °F to 140 °F (-20 °C to 60 °C),
<95% RH
N/A
IP 40
N/A
N/A
N/A
N/A
N/A
1.26 in (32 mm)
Approx. 5.9 x 2.75 x 1.18 in
(150 x 70 x 30 mm)
Approx. 0.25 lb (0.114 kg)