Measurement Initialization Error; No Configuration File Loaded; Selected File Is Incompatible; Slow Or Missing Clock - HP E2470A Manual

Preprocessor interface for motorola mc68hc16y1
Table of Contents

Advertisement

"Measurement Initialization Error"

This error occurs when you have installed the cables incorrectly for one or
two HP 16550A logic analysis cards. Ensure that your cable connections
match the silk screening on the card. Then, repeat the measurement.
See Also
The HP 16550A 100-MHz State/500-MHz Timing Logic Analyzer Service
Guide.

"No Configuration File Loaded"

This is usually caused by trying to load a configuration file for one type of
module/system into a different type of module/system.
Verify that the appropriate module has been selected from the Load
{module} from File {filename} in the HP 16500A/B disk operation menu.
Selecting Load {All} will cause incorrect operation when loading most
preprocessor interface configuration files.
See Also
Chapter 1 describes how to load configuration files.

"Selected File is Incompatible"

This occurs when you try to load a configuration file for the wrong module.
Ensure that you are loading the appropriate configuration file for your logic
analyzer.

"Slow or Missing Clock"

This error message might occur if the logic analyzer cards are not firmly
seated in the HP 16500A/B or HP 16501A frame. Ensure that the cards
are firmly seated.
This error might occur if the target system is not running properly.
Ensure that the target system is on and operating properly.
If You Have a Problem
"Measurement Initialization Error"
A–11

Advertisement

Table of Contents
loading

Table of Contents